Mts Landmark Testing Systems Manual de usuario Pagina 19

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 34
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 18
Hydraulic Power Unit Specifications
Landmark Test System Site Prep Guide System Component Specifications
19
Model Specific Specifications
Parameter 505.07 Specification 505.11 Specification
Pump/motor
Maximum Continuous
pressure
Maximum Flow
capacity
Motor rating
Line voltage starter configuration
21 MPa (3000 psi)
22.7 L/m (6 gpm) at 50 Hz
26.5 L/m (7 gpm) at 60 Hz
11.2 Kw (15 hp) at 50/60 Hz
Wye-Delta starter configuration
21 MPa (3000 psi)
41.6 L/m (11 gpm) at 50/60 Hz
18.6 Kw (25 hp) at 50/60 Hz
Water flow rating
(input temperature)
10.0°C (50°F)
15.5°C (60°F)
21.1°C (70°F)
26.7°C (80°F)
32.2°C (90°F)
Heat load (maximum)
3.8 L/m (1.0 gpm)
4.9 L/m (1.3 gpm)
6.1 L/m (1.6 gpm)
8.3 L/m (2.2 gpm)
15.9 L/m (4.2 gpm)
12.3 kW (42,000 Btu/hr)
7.2 L/m (1.9 gpm)
9.1 L/m (2.4 gpm)
12.1 L/m (3.2 gpm)
18.9 L/m (5.0 gpm)
49.2 L/m (13.0 gpm)
20.5 kW (70,000 Btu/hr)
Weight with maximum oil
174 L (46 gal))
450 kg (992 lb) 473 kg (1042 lb)
Vista de pagina 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 33 34

Comentarios a estos manuales

Sin comentarios