Mts Landmark Testing Systems Manual de usuario Pagina 24

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 34
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 23
Landmark Test System Site Prep Guide
24
Controller Specifications
System Component Specifications
Specifications–Model 494.06 Chassis
All equipment related to the controller should be connected to the same fused
power circuit.
Note Electrical connections must be made by qualified personnel and conform
to local codes and regulations. Local electrical codes supersede any
information found here.
Model 494.06 Chassis Specifications
Parameter Specification
Input Voltage 100–240 V AC (single phase)
power factor corrected universal input
Input Frequency 50–60 Hz
Input Current 10 A at 100 V AC
5 A at 240 V AC
Facility Power Requirements Provide a dedicated circuit for the
chassis, computer, and monitor.
Input Surge <40 A
Insulation Over Voltage Category I
Pollution Degree 2
Weight Approximately 14 kg (31 lb)
Dimensions Width: 21.6 cm (8.5 in)
Height: 44.2 cm (17.4 in)
Depth: 64.8 cm (25.5 in)
Vista de pagina 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 33 34

Comentarios a estos manuales

Sin comentarios