5. Reset the interlocks if needed and select the Low, then High, power buttons.
6. Click Run to run the test.
5.4.2.0 Set the FGC DCPD Test Parameters
5.4.2.1.0 Test Parameters Overview
After a new test run is added to the test, the Main Menu window appears.This window provides access to
all necessary test parameters. Click the parameter buttons in the Main Menu window to specify parameter
values for the test.
5.4.2.2.0 FCG DCPD Parameter Values for Simulation
The sample test includes a test run with test data that can be analyzed. If you want to reproduce the same
results in another test run, add another test run, and then specify the following values for the test parameters.
Other parameters not listed here should remain unchanged.
Precrack Parameters
Precrack Parameter Values
Value to Use for SimulationParameter
10 HzPrecrack Frequency
25 mmPrecrack Final Crack Limit
1 kNPrecrack Measure Load Level
0.1 unitlessPrecrack Load Ratio
0.1 kN/mm^1.5Precrack Final Maximum K
1000000 countPrecrack Cycle Limit
Crack Growth Parameters
Delta K Parameter Values
Value to Use for SimulationParameter
True SineFCG Wave Shape
10 HzFCG Frequency
0.2 kN/mm^1.5FCG Delta-K Initial
0.1 unitlessFCG Load Ratio
MTS TestSuite 57
5.0 Running Simulation Tests