Edit the Calibration File
SWIFT 10 ATV Sensors Setting up the Transducer Interface
57
Edit the Calibration File
Do not change any other items in the calibration file.
The calibration file contains offset values for all of the bridge outputs.
Changing any of the items other than those listed in the following procedure
will cause your calibration file to be incorrect.
Take care not to change any values except those listed in the following procedure.
If your calibration file is incorrectly changed, reload the original file from the
diskette provided by MTS.
1. Open the calibration file using a text editor.
Typically two calibration files are provided with each transducer: one for
spinning applications and one for non-spinning (fixed) applications.
The spinning application files are usually names in the format <transducer
serial number>s.cal.
The non-spinning (fixed) application files are usually names in the format
<transducer serial number>f.cal.