Mts Series 793 Controller Manual de usuario Pagina 90

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 106
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 89
MTS Series 793 Controller Overview
Steps to Run a Test
Perform a Practice Test
90
C. Select the following settings in the Function Generator.
D. From the Station Manager application toolbar, select the Scope.
A crush zone exists between the actuator and any equipment in the path of
its movement.
A crush zone exists between the actuator and any equipment in the path of
its movement. Immediate and unexpected actuator response is possible
when you apply hydraulic pressure to your system.
Stay clear of the actuators when applying hydraulic pressure. Do not place any
part of your body in the path of a moving actuator.
CONTROL SETTING
Channel
Channel 1
Control Mode
Force
Command Type
Cyclic
Target Setpoint
0 kN
Amplitude
10% of full scale
Frequency
1 Hz
Wave Shape
Ramp
Compensator
None
WARNING