SWIFT 40 Sensors
28
Specifications
Hardware Overview
Transducer Interface Communications
P
ARAMETER SPECIFICATION
Communications Channel # 1
(Remote Host Connections)
Baud rates
Parity
Stop bits
Data bits
Isolated RS-232/RS-485 interface power
supply
Electrical interface
Maximum number of devices that can be part
of a RS-485 multidrop chain
19,200 Kbits/s
None
1
8
+5 V DC @ 200 mA maximum
Isolated RS-232 or RS-485 remote host connection
Isolated RS-485 TI to TI connection
32 with RS-232 remote host
*
31 with RS-485 remote host
* Includes all compatible devices, such as an MTS 407 Controller. A maximum of only nine transducer
interfaces can be connected, because the addresses are limited to 1–9.
Maximum cable length
For RS-232 host:
50 ft from host to the first (nearest) TI,
and
300 ft from the first TI to the last SWIFT TI in the RS-
485 multidrop chain
For RS-485 host:
300 ft from host to the last (furthest) TI in the RS-485
multidrop chain