Set up a Scope to compare the command signal to the resulting feedback. Enlarge the scope if
necessary, and then set up the scope as follows:
Y1 parameters:
Channel Ch1 (or other depending on your configuration)
Signal Command
Unit mm
Units/div 1.0000
Offset 0.0000
Y2 parameters:
Channel Ch1 (or other depending on your configuration)
Signal Displacement
Unit mm
Units/div 1.0000
Offset 0.0000
Time:
Trace Time — 5.0000 s
5. Set the HPU and HSM to Power High.
68 | MTS TestLine Test System Operation
Running the Example HCF Test