Because the specimen is fixed solidly in place, you can set narrow displacement limits that will trip if
anything breaks.
a) Click the Station Setup button.
b) Select Channels > Ch1 > Displacement > Channel Input Signals button > Limits tab.
c) Set the Upper Limit to 5 mm.
d) Set the Upper Action to Interlock.
e) Set the Lower Limit to -5 mm.
f) Set the Lower Action to Interlock.
2. Set force limits that are just outside the expected values for tuning.
MTS TestLine Test System Operation | 85
Running the Example HCF Test