MTS FlexTest® Models 40/60/100/200 Controller Hardware
Model 494.33 Digital I/O Power Supply
Digital I/O and Transition Boards
228
Specifications–Model 494.33 DI/O Power Supply
Electrical
AC power disconnect To disconnect AC power from the power supply, unplug the AC power cable
from the front of the power supply.
P
ARAMETER SPECIFICATION
Input Voltage 100–240 V AC, 5–2 A, single phase
“J1 Output Pwr”
Output
24 V DC, 14 A maximum (7 A per output)
*
Fuse: 7 A, 250 V, Slo-Blo
* The maximum cable length for all power cables is 6 m (20 ft).
When used in an electrical circuit, the voltage and current (amperage)
ratings of a fuse are selected to limit the capacity of the circuit and to
provide protection of the devices (equipment) in the circuit.
Using fuses with ratings that are different from the ones specified can
result in fire or electrical damage to the equipment. Fire can also result
in injury or death to personnel.
Before replacing a fuse, troubleshoot and correct the cause. Look for
damaged conductors (for example, wires, switches, relays, terminals, and so
on) and faulty equipment. Replace a fuse only with one that has the same
ratings.
“J3 Input Pwr”
Output
24 V DC, 1 A maximum*
Fuse:1 A, 250 V, Slo-Blo
WARNING