Installation
MTS FlexTest® Models 40/60/100/200 Controller Hardware FlexTest Controller Configurations
35
Environmental Requirements—Series 494 Hardware
All Series 494 hardware components are intended for indoor use only. This
indoor environment must conform to the following environmental specifications.
Note All Series 494 Controller must only be operated under the installation
and ambient conditions (such as, temperature, moisture, and EMC)
specified.
Note To maintain EMC compliance, the controller must be installed in a
location that does not exceed the EN 61000-6-4 emission standard for
industrial environments.
PARAMETER SPECIFICATION
Temperature 5ºC–40ºC (41ºF–104ºF)
Humidity 5–85%, non-condensing
Altitude 3048 m (10,000 ft) maximum
Space
Requirements
For proper ventilation, allow 51 mm (2 in) clearance
on all sides of the chassis.
The rear of the chassis requires a minimum
clearance of 15.24 cm (6 in) for cable connections.