Mts FlexTest Models 200 Manual de usuario Pagina 87

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 344
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 86
Model 494.44 Two-Station System Board
MTS FlexTest® Models 40/60/100/200 Controller Hardware FlexTest Controller Configurations
87
Auxiliary Power
Outputs
+5 V DC
+15 V DC
-15 V DC
+24 V DC
Connector J49 (D9S)
0.75 A maximum
0.75 A maximum
0.75 A maximum
0.75 A maximum
Digital Outputs Connector J55 (D9S)
Note Digital output signals must be routed
through a Model 494.32 DI/O Breakout Box
that includes drivers and connections for
external devices.
Digital Inputs Connector J54 (D9P)
Note Digital input signals must be routed through
a Model 494.32 DI/O Breakout Box that
includes opto-isolators and connections for
external devices.
D/A Outputs
Resolution
Output Type
Output Voltage
Output Current
Connector "D/A Output" (D9P)
16 bit
Single ended
+/- 10 V full scale
5.0 mA maximum
Note This connector is located on the rear panel
of the Model 494.04 Chassis.
Model 494.44 Specifications (part 3 of 3)
P
ARAMETER SPECIFICATION
Vista de pagina 86
1 2 ... 82 83 84 85 86 87 88 89 90 91 92 ... 343 344

Comentarios a estos manuales

Sin comentarios