SWIFT 20 Sensors
48
TIXFER - Low-Profile Transducer Interface Transfer
Software Utilities
TIXFER - Low-Profile Transducer Interface Transfer
This program is used to change settings within the SWIFT Low Profile
Transducer Interface (TI). It can be used to read the current settings and save
them to the computer (upload) or write the settings to a Low-Profile TI with
values from a file on the computer (download).
Syntax The tixfer command requires one argument:
tixfer port
port is the serial port number
1 = COM1:
2 = COM2:
Example
File Format The file used with commands contains a header and a list of parameters.
Transducer calibrations can be uploaded and saved individually. If a test needs to
be rerun at a later date but the original transducer is not available, another
transducer can be used by downloading its calibration information.
The header must be the first thing in the file, and it must be in the form:
[SWIFT]
Following the header is a list of parameter settings. The syntax is:
ParamName=ParamValue
C:\bin>tixfer 1
tixfer $Revision: 1.11 $ ($Date: 1999/04/21 10:12:40 $)
0...exit
1...Upload/Save a chassis
2...Restore/Download a chassis
Enter choice? 1
Enter TI box address? 1
Enter output specifications filename? example.cal
Enter description? example for uploading settings
Allocating box...
Initializing box...