HSM Connections
Models FlexTest® IIm/GT/SE Controller Hardware FlexTest IIm Controller Connections
191
HSM Connections
The outputs provide 24 V DC to turn on the HSM solenoids.
Cable specification P28 is a 4-contact CPC male connector (AMP Incorporated).
Cable—18 AWG/2 connector type SJOOW-A for on/off HSMs 18 AWG/3
connector type SJOOW-A for high/low HSMs
Cable assembly numbers:
290.XX HSM, on/off, 24 V DC, PN 039-705-3XX 18/2 SJO
290.XX/294.XX HSM, high/low, 24 V DC, PN 039-705-5XX 18/3
SJO
298.11 HSM, on/off, 24 V DC, PN 039-708-1XX 18/2 SJO
1
2
4
Low Pressure
DC Common
High Pressure
To HSM
Solenoids
J28From
Hydraulic I/O
Connector J28 controls the
low-pressure and high-
pressure solenoids of a
hydraulic service manifold.