Mts Model FlexTest SE Manual de usuario Pagina 92

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 264
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 91
Models FlexTest® IIm/GT/SE Controller Hardware
Model 493.21B (DUC B) Jumpers
FlexTest GT Controller Connections
92
Force/strain sensor
connections
Force and strain sensors require a DC conditioning mezzanine card.
Model 493.21B (DUC B) Jumpers
The following jumpers are for the Model 493.21B Digital Universal
Conditioner). The jumpers are labeled W instead of X; the W jumpers are solder
pads. The following jumper descriptions are from the 493.XX Module Setup
configuration drawing (PN 049-336-301).
Active guard Jumper W1 enables or disables an active guard for the sensor cable.
Active guard is usually not required. In particular environments, active guard
drive may reduce transducer noise.
4
12
2
15
1
10
7
3
5
13
9
8
11
14
6
E4
E2
E1
E7
E3
E5
Inner
Guard
4
12
2
15
1
10
3
5
13
9
8
11
14
6
Remote Cal
Clk
Data to ID
Data from ID
Sig Gnd
Transducer
ID Module
ID Circuit
Remote Cal
-FB
-FBR
+FB
+FBR
-EX
-EXS
+EX
+EXS
DC Sensor
Backshell
GND
Future
Future
Future
Future
W1
W1
Enable Disable
Default
Vista de pagina 91