Digital Universal Conditioner (DUC BC) Switches
Models FlexTest® IIm/GT/SE Controller Hardware FlexTest GT Controller Connections
95
4- or 8-Wire
sensor cable
Switches S1-1, S1-2, S1-3, and S1-4 establish the correct shunt calibration
connections for 4-wire and 8-wire sensor cables. The length of the sensor cable
determines which type of cable is used. Short cables are typically 4-wire cables
while long cables are typically 8-wire cables.
Switches S1-1 and S1-2 select a 4- or 8-wire transducer for positive
feedback.
Switches S1-3 and S1-4 select a 4- or 8-wire transducer for negative
feedback.
The setting of jumper W8 (single ended excitation enable) may affect your
choices for these switches.
Note +/– excitation (–excitation is standard).
Sensor Cable Settings (4/8-Wire)
REQUIRED CONNECTION SWITCH SETTINGS
-EX to -FB shunt in 4-wire
configuration
DEFAULT
S1-1 OFF
S1-2 OFF
S1-3 ON
S1-4 OFF
-EX to +FB shunt in 4-wire
configuration
S1-1 ON
S1-2 OFF
S1-3 OFF
S1-4 OFF
-EX to -FBR shunt in 8-wire
configuration
S1-1 OFF
S1-2 OFF
S1-3 OFF
S1-4 ON
-EX to +FBR shunt in 8-wire
configuration
S1-1 OFF
S1-2 OFF
S1-3 ON
S1-4 OFF